Some users may have more than one Dimension Icon scanner for their system. For the software to function correctly, it must be able to identify the scanner in use. Use the NanoScope 8.15 functionality to correctly swap the scanner.
|
|
|
|
(Hover over the image to view larger)
|
(Hover over the image to view larger)
|
|
NOTE: The Next button will not be activated until you move the stage to a safe position using the Move Stage button.
NOTE: You may not see this screen if the stage is already in the safe position.
|
|
(Hover over the image to view larger)
|
To remove the Dimension Icon scanner:
|
(Hover over the image to view larger)
|
To connect the Dimension Icon scanner:
NOTE: Dovetail engagement actuates with a spring. Failing to engage the spring-loaded dovetail causes a large increase in image noise due to reduced rigidity of the mechanical support of the scanner.
|
|
|
Click OK and wait while the system writes the scanner file to the appropriate location. When you enter Realtime the System Status Indicator should display no problems, indicating that the software is now using the correct scanner file.
|
| www.bruker.com | Bruker Corporation |
| www.brukerafmprobes.com | 112 Robin Hill Rd. |
| nanoscaleworld.bruker-axs.com/nanoscaleworld/ | Santa Barbara, CA 93117 |
| Customer Support: (800) 873-9750 | |
| Copyright 2010, 2011. All Rights Reserved. |